Cup and Head Combination Options
The Tesera Trabecular Acetabular System includes configurations for head sizes up to 40 mm and cup sizes up to 66 mm. Femoral heads are available in CoCr and BIOLOX®† delta ceramic.
Components with the Tesera structure are created using electron beam manufacturing (EBM). The solid and porous regions of the components are built up layer by layer from titanium-alloy powder in one continuous process; thus, the Tesera structure is integral to the component and is not a coating.
Cup Sizes (mm) | 44 | 46 | 48 | 50 | 52 | 54 | 56 | 58 | 60 | 62 | 64 | 66 | |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Insert Sizes (mm) | 44/46 | 48/50 | 52/54 | 56/58 | 60/62 | 64/66 | |||||||
Head Sizes (mm) | 28 | X | X | X | X | ||||||||
32 | X | X | X | X | |||||||||
36 | X | X | X | X | X | X | X | X | |||||
40 | X | X | X | X | X | X |
Insert Specifications
All inserts are available in both MAX Highly Crosslinked Polyethylene and E-MAX Highly Crosslinked Polyethylene.
Insert Size | Inner diameter (mm) | Thickness (at dome) (mm) | Hood Height (mm) |
---|---|---|---|
44/46 | 28mm | 4.9 | 3.3 |
48/50 | 28mm | 6.8 | 3.7 |
48/50 | 32mm | 4.8 | 3.7 |
52/54 | 32mm | 6.8 | 4 |
52/54 | 36mm | 4.8 | 4 |
56/58 | 36mm | 6.8 | 4.4 |
60/62 | 36mm | 8.8 | 4.7 |
64/66 | 36mm | 10.8 | 5.1 |
56/58 | 36mm or 40mm | 4.8 | 4.4 |
60/62 | 36mm or 40mm | 6.8 | 4.7 |
64/66 | 36mm or 40mm | 8.8 | 5.1 |
Femoral head options
Size (mm) | Offset (mm) | |||||
---|---|---|---|---|---|---|
CoCr Alloy | 28 | -4 | 0 | +4 | +8 | +12 |
32 | -4 | 0 | +4 | +8 | +12 | |
36 | -4 | 0 | +4 | +8 | +12 | |
40 | -4 | 0 | +4 | +8 | +12 | |
BIOLOX delta Ceramic | 28 | -3.5 | 0 | +3.5 | – | – |
32 | -4 | 0 | +4 | +7 | – | |
36 | -4 | 0 | +4 | +8 | – | |
40 | -4 | 0 | +4 | +8 | – |
†BIOLOX delta is a registered trademark of Cerasiv GmbH Innovatives Keramick-Engineering and CeramTec AG Innovative Ceramic Engineering.